Description
FDA RFQ 75F40126Q134116
This is a combined synopsis/solicitation for commercial products or commercial services prepared
in accordance with the Federal Acquisition Regulation (FAR) RFO Part 12. This announcement
constitutes the only solicitation. Offers are being requested and a separate written solicitation will not
be issued. The solicitation number is 75F40126Q134116. This solicitation is issued as a Request for
Quote (RFQ) for Preventive maintenance-repair (PM) service agreement for Leica CARS SP8
confocal microscope.
This acquisition is not set-aside for small business concerns. This solicitation incorporates provisions
and clauses by reference. The full text of provisions and clauses may be accessed electronically at
https://www.acquisition.gov/far-overhaul.
The associated North American Industry Classification System (NAICS) code for this acquisition is
811210 – Electronic and Precision Equipment Repair and Maintenance, with a Small Business Size
Standard of $34 million. Any business, regardless of size, may respond with a quote to this solicitation.
Schedule of Supplies or Services and Prices/Costs
Services shall be performed at the U.S. FDA Center for Biologics Evaluation and Research Microscopy
Lab, Bldg. 52/72, Room 1108, 10903 New Hampshire Avenue, Silver Spring, MD 20993-0002.
Line Item 01 Base Period
Preventive maintenance-repair (PM) service agreement for Leica CARS SP8 confocal
microscope S/N: 8100000945; to include CARS Laser S/N: S06655, UC7_FC7 Basic
Instrument S/N: 602425, and DMi8 Microscope Uni f/Mot, S/N: 416020, and
meeting the minimum technical requirements set forth below:
Quantity: 1
Unit: 12 consecutive months
Annual Price: ____________________
Period of Performance: 7/1/2026 – 6/30/2027
GSA Contract Number (if applicable): ____________________________
Line Item 02 Option Year 01
Preventive maintenance-repair (PM) service agreement for Leica CARS SP8 confocal
microscope S/N: 8100000945; to include CARS Laser S/N: S06655, UC7_FC7 Basic
Instrument S/N: 602425, and DMi8 Microscope Uni f/Mot, S/N: 416020, and
meeting the minimum technical requirements set forth below:
Quantity: 1
Unit: 12 consecutive months
Annual Price: ____________________
Period of Performance: 7/1/2027 – 6/30/2028
GSA Contract Number (if applicable): ____________________________
-- 1 of 14 --
Line Item 03 Option Year 02
Preventive maintenance-repair (PM) service agreement for Leica CARS SP8 confocal
microscope S/N: 8100000945; to include CARS Laser S/N: S06655, UC7_FC7 Basic
Instrument S/N: 602425, and DMi8 Microscope Uni f/Mot, S/N: 416020, and
meeting the minimum technical requirements set forth below:
Quantity: 1
Unit: 12 consecutive months
Annual Price: ____________________
Period of Performance: 7/1/2028 – 6/30/2029
GSA Contract Number (if applicable): ____________________________
Background and Requirement
The U.S. Food and Drug Administration (FDA), Center for Biologics Evaluation and Research
(CBER), Microscopy Lab requires a preventive maintenance and repair agreement for its Leica CARS
SP8 confocal microscope S/N: 8100000945; to include CARS Laser S/N: S06655, UC7_FC7 Basic
Instrument S/N: 602425, and DMi8 Microscope Uni f/Mot, S/N: 416020.
This microscope is heavily used by CBER and CDER researchers across the FDA’s White Oak
campus in Silver Spring, MD, and is used for live and fixed fluorescence and non-linear microscopy.
The instrument is currently not covered under an existing agreement.
Minimum Technical Requirements are as follows for preventive maintenance and repair
service for a Leica CARS SP8 confocal microscope S/N: 8100000945; to include CARS Laser
S/N: S06655, UC7_FC7 Basic Instrument S/N: 602425, and DMi8 Microscope Uni f/Mot,
S/N: 416020:
*One (1) planned annual preventive maintenance (PM) visit for the SP8 AOBS IR system, (Premium
Care Service Plan coverage), to include labor, travel and parts to ensure the system is performing
optimally and to factory specifications;
*One (1) planned annual PM visit for the EM UC7 ultramicrotome (Advanced Care Service Plan
coverage;
*PM’s shall cover the entire microscope system - microscope stand, objective lens, any accessories as
well as attached laser systems including UV and Visible and APE picoEmerald CARS IR Lasers and
associated chiller, scan head, detectors, DMi8, UC-7, Tokai Hit stage top incubator for live-cell
imaging and LIS large environmental chamber and attachment, automated stage controller and any
microscope related parts, and any accessories and instrument control workstation and software;
*Shall include unlimited service visits by Leica-Microsystems trained and qualified field service
engineers;
*Shall include unlimited replacement service parts, components, or modules necessary for repairs,
excluding consumables, 3rd Party parts, and IR/MP lasers
*All maintenance and repair activities shall be performed only by factory-trained and certified Leica
field service representatives, following OEM specifications, manuals, and service bulletins, using
OEM replacement parts, components, subassemblies, etc.;
*In the event that a laser or other system component fails and cannot be repaired on-site, within 5
days the contractor shall provide loaner parts and shipping materials/packaging until the FDA parts
have been repaired and replaced. without limitation of loaner time, within five business days;
*24-hour e-mail or telephone response to a reported instrument issue; 72-hour response time for a
Leica field engineer to be on-site for repairs;
-- 2 of 14 --
*Prioritized on-site response time for remedial services (critical services);
*Unlimited access to remote support Monday through Friday during normal business hours (8:00 am
– 5:00 pm local time, excluding Federal holidays);
*Shall include software and firmware updates as required by the Original Equipment Manufacturer
(OEM) either remotely or in conjunction with a scheduled preventive maintenance or corrective
maintenance visit (updates are defined as changes of t…