NASA is seeking to procure a Scanivalve Calibration System, specifically a Scanivalve SPC4050 with various components, including a pneumatic logic unit and a portable calibration system. This is a sole source requirement under NAICS 334513, suitable for small businesses that can provide the specified calibration equipment. The notice type is a Combined Synopsis/Solicitation, indicating that interested vendors should prepare to respond to the solicitation rather than just track it.
Analysis by Mindy, grounded in the SAM.gov notice.
Description
STATEMENT OF WORK
1. Objective/Requirements
The NASA Glenn Research Center requires the purchase of a Scanivalve Pressure
Calibrator.
2. Characteristics, Scope, and Specs
LINE PART NO. DESCRIPTION QTY.
1 SPC4050 SPC4050 2-Channel Pressure Calibrator
(SPECIAL: with
IS-33, 0.008% FS Accuracy), with internal
barometer
Channel A: Low Pressure Servo Regulator
(LPSVR)
Transducer 1: 5psig (Cal'd 5.75psig)
Transducer 2: 15psig (Cal'd 17psig)
Channel B: Medium Pressure Servo
Regulator (MPSVR)
Transducer 1: 50psig (Cal'd 56psig)
Transducer 2: 100psig (Cal'd 115psig)
1
2 SPC4050PLU-2 Pneumatic Logic Unit, 2-Channel (90psi
CTL)
1
3 70380-02 Pneumatic Distribution Panel, SPC4050, 2
Channel, 90 psi Control
1
4 73680 4-Range Portable Calibration System with
items 8-10 installed. Including Pneumatic /
Electrical interface panel, internal
pneumatic and electrical connections,
testing and documentation. REF: Sales
Dwg 80068-01
1
System shall support the following pressure controller/ calibrator:
Scanivalve SPC4050 with the following salient specifications:
• ChA and ChB transducers - IS-33, 0.008% FS Accuracy
• Internal barometer channel
• Channel A: Low Pressure Servo Regulator (LPSVR) Transducer 1: 5psig (Cal
5.75psig)
• Transducer 2: 15psig (Cal 17psig)
• Channel B: Medium Pressure Servo Regulator (MPSVR) Transducer 1: 50psig (Cal
56psig)
• Transducer 2: 100psig (Cal 115psig)
• Removeable/ Interchangeable transducers
-- 1 of 2 --
• Color touch screen
• Used for automated DSA5002 calibrations
• Bench or rack-mount
System shall support the following pneumatic logic unit:
Scanivalve SPC4050PLU-2 Pneumatic Logic Unit with the following salient specifications:
• 2 – channel, 90 psi control pressure
• Positive or below atmospheric (vacuum) calibrations
• Zero offset correction logic
• Leak and response testing on pressure scanners
System shall support the following pneumatic distribution panel:
Scanivalve 70380-02 Pneumatic Distribution Panel with the following salient specifications:
• 2 – channel, 90 psi control pressure
• Provides all necessary supply and control pressure to SPC4050 and SPC4050PLU-2
System shall support the following portable calibration unit/ case:
Scanivalve 73680 4-range Portable Calibration System with the following salient specifications:
• SPC4050 installed
• SPC4050PLU-2 installed
• 70380-02 installed
• Durable case for the above items to allow for ease of transport
• Protection for sensitive instrumentation in transit
Warranty:
The vendor shall provide a warranty against defects in materials and workmanship for a
period of 1 year following completion of work. Any nonconformities identified by NASA as
defects in materials and workmanship from the original delivery shall be handled as Non-
Conforming Deliverables as described above.
3. Place of Performance
NASA Glenn Research Center
21000 Brookpark Rd
Cleveland OH 44135-3127
4. Period of Performance
14 weeks ARO.
-- 2 of 2 --
Source: SAM.gov, as posted. Verify the current solicitation before responding.