The Department of Defense is seeking a Total Small Business Set-Aside contract for an on-site, fully automated Liquid Nitrogen Generator that meets or exceeds the technical and functional capabilities of the Rutherford & Titan - Titan 10 L/Day Liquid Nitrogen Generator. The system must meet specific performance and physical parameters, including production capacity, nitrogen purity, and environmental tolerance, and possess automated capabilities such as intelligent production control and remote monitoring and diagnostics. This opportunity is suitable for small businesses that manufacture equipment in the NAICS code 336413, and bidders should track the Combined Synopsis/Solicitation notice.
Analysis by Mindy, grounded in the SAM.gov notice.
Description
SECTION 1: Scope of Requirement
The Department of War has a requirement for an on-site, fully automated Liquid Nitrogen
(LN2) Generator. The system must meet or exceed the technical and functional capabilities
of the Rutherford & Titan - Titan 10 L/Day Liquid Nitrogen Generator (Brand Name or Equal).
SECTION 2: Salient Characteristics (Brand Name or Equal)
To be considered technically acceptable, any proposed "or equal" system must meet or
exceed the following objective performance and physical parameters:
Characteristic Requirement / Minimum Specification
Production Capacity Not to exceed 10 Liters per Day (L/Day).
Integrated Storage
Capacity
Not to exceed 55 Liters (55L) internal storage tank.
Nitrogen Purity Greater than 99% pure liquid nitrogen.
Physical Dimensions Maximum footprint of 63″ H × 25″ D × 36″ W to fit existing spatial
constraints.
System Weight Maximum of 450 lbs.
Mobility Must be integrated with heavy-duty rollers/casters for on-site mobility.
Cooling Method Air-cooled system (no external water-cooling infrastructure required).
Power Requirements Must operate standard electrical configurations: 120V, 60Hz, maximum
draw of 20 Amps.
Power Consumption Energy efficient; maximum of 2.2 kW/H during active production.
Acoustic Signature Maximum noise level of 60–65 dBA at a distance of 2 feet.
Environmental
Tolerance
Must be capable of continuous, automated operation in high ambient
temperatures up to +125°F (+51.6°C).
SECTION 3: Key Functional & Smart Features
Any proposed "or equal" system must possess equivalent automated capabilities as
described below:
Feature Performance and Functional Standards
Automated
Operation
"Plug-and-play" capability; fully automated startup/shutdown cycles requiring
only power connection and a single manual/digital "start" input.
-- 1 of 2 --
Intelligent
Production
Control
Must feature an integrated, intelligent control system (equivalent to FlowSync™
Technology) that continuously monitors internal LN₂ levels to automatically
adjust, start, and stop production based on usage and capacity.
Remote
Monitoring &
Diagnostics
Must include built-in remote access capability (via Wi-Fi or Ethernet) to allow
real-time system monitoring, troubleshooting, and diagnostics (equivalent
to SmartConnect™ Monitoring).
Security
Standards
Any wireless/wired connectivity must be compliant with standard federal
networks or capable of being completely disabled/isolated based on local
security protocols.
-- 2 of 2 --
Source: SAM.gov, as posted. Verify the current solicitation before responding.