The Department of Commerce is seeking a CMOS Microelectrode Array System, classified under NAICS code 334513, with a primary place of performance in Maryland. This solicitation is suitable for small businesses, particularly those with expertise in microelectrode array systems and related technologies. The notice type is a Combined Synopsis/Solicitation, indicating that vendors should track the solicitation for updates rather than submitting a bid immediately, and the solicitation requires synchronization and stimulus delivery capabilities.
Analysis by Mindy, grounded in the SAM.gov notice.
Description
Amendment 0001:
The purpose of this amendment is to answer questions submitted by vendors and extend the solicitation response date.
Question 1: What is your recording subject?
NIST Answer: As listed in Statement of Work (SOW), the measurement subjects are in vitro neurons, neuronal organoids, and micro physiological systems with various cell types such as neurons or cardiomyocytes.
Question 2: How many recording channels channel are you looking for? This will determine the channel capacity needed and head stages to be used.
NIST Answer: As specified in the SOW, ≥ 1000 simultaneous Voltage recording channels per well at minimum.
Question 3: What is your implant type and connector?
NIST Answer: As specified, the in vitro cells will be cultured directly on top of the HD-MEA surface and a culture well to contain the media is required.
Question 4: Is the subject going to be head-fixed. anesthetized, or freely moving during the recording.
NIST Answer: No, the cells will be cultured directly on top of the HD-MEA surface.
Question 5: For freely-moving wired subjects, we also offer a motorized commutator called the Saturn. Are you interested in commutator?
NIST Answer: No, the cell location will be fixed directly on top of the HD-MEA surface.
Question 6: Do you require external device synchronization or planning to deliver a stimulus?
NIST Answer: Yes, both are required. From the specification in SOW: Hardware is required for synchronizing and/or triggering recordings to external equipment, and a software API is required for active control over experiments, including closed-loop feedback experiments and automation of functions using sequencing and/or triggering.
Question 6: Do you need video tracking camera setup?
NIST Answer: No, NIST has existing microscopy imaging systems that would be suitable for the specified HD-MEA system.