The Department of Veterans Affairs is buying testing and reporting services for utility and critical water systems at the Asheville VAMC, which suits businesses with a NAICS code of 541330. This is a solicitation notice, meaning businesses should track the opportunity to prepare for a potential bid. The work involves testing water quality to meet specific standards, with testing to occur at the Sterile Processing Service department in Asheville, NC.
Analysis by Mindy, grounded in the SAM.gov notice.
Description
B.3 PERFORMANCE WORK SCHEDULE
1. Scope of Work
a. The scope of this work is all labor and materials to provide testing and reporting services per ANSI/AAMI ST108: 2023 for utility and critical water systems. The goal is for the successful bidder to test the critical water system to prove compliance with Table 1:
Type of Water
Utility Water(1)
Critical Water
Water Use
Units
Flushing/Washing/Rinsing
Final Rinse(2)/ Steam
pH(4)
pH
6.5 – 9.5
5.0 – 7.5
Conductivity (mg/L = ppm)
µS/cm
< 500
< 10
Total Alkalinity
mg CaCO3
/L
< 400
<8
Total Hardness
mg CaCO3
/L
< 150(3)
< 1
Bacteria
CFU/mL
< 500
< 10
< 10
Endotoxin
EU/mL
N/A
< 10
< 10
Total Organic Carbon (TOC)
mg/L
N/A
N/A
< 1.0
Color and Turbidity
Visual
Colorless, clear, without sediment
Colorless, clear, without sediment
Colorless, clear, without sediment
NOTE 1: This is the quality of water that might come from the tap but might need some form of treatment to achieve these specifications.
NOTE 2: If this is the final rinse prior to sterilization of a critical device.
NOTE 3: If hardness is greater than 150 mg/L, a water softener is recommended unless used for washing and the cleaning chemistry is capable of handling higher levels of hardness.
NOTE 4: For boiler-treated steam, most boilers are treated to maintain a pH of 7.5 or 9.2. Any treatment of water that goes into boilers should be in accordance with the sterilizer and boiler manufacturers’ written IFU.
Table 1: Categories and recommended levels of water quality for medical device reprocessing.
2. Location of Work
a. Testing services shall occur at the Sterile Processing Service department, located in the basement of Building 47 at 1100 Tunnel Road, Asheville, NC, 28803.
3. Testing
Testing shall be provided within the Table 2 schedule below:
Characteristic
Type of testing
Sample site
Samples taken and analyzed by
Suggested frequency of testing
CGVAMC Responsible Service
Bacteria
Heterotrophic plate count (1)
Reprocessing area at each of the 11 point-of-use sites (3 - critical sinks, 1 - utility sink, 1 - RO storage tank, 4 - medivator sterilizers, 2 – washers). Immediately downstream of water treatment process.
Contractor
Weekly
FMS (contract)
Endotoxins
LAL test
Reprocessing area at each of the 11 point-of-use sites (3 - critical sinks, 1 - utility sink, 1 - RO storage tank, 4 - medivator sterilizers, 2 – washers). Immediately downstream of water treatment process.
Contractor
Monthly
FMS (contract)
pH
pH meter
Reprocessing area at each of the 11 point-of-use sites (3 - critical sinks, 1 - utility sink, 1 - RO storage tank, 4 - medivator sterilizers, 2 – washers). Immediately downstream of water treatment process.
Contractor
Monthly
FMS (contract)
Total hardness
Determination of ppm CaCO3
by handheld meter, titration kit, or colorimetric dipsticks.
Reprocessing area at each of the 11 point-of-use sites (3 - critical sinks, 1 - utility sink, 1 - RO storage tank, 4 - medivator sterilizers, 2 – washers). Immediately downstream of water treatment process.
Contractor
Monthly
FMS (contract)
Conductivity
Conductivity Meter
Reprocessing area at each of the 11 point-of-use sites (3 - critical sinks, 1 - utility sink, 1 - RO storage tank, 4 - medivator sterilizers, 2 – washers). Immediately downstream of water treatment process.
Contractor
Monthly
FMS (contract)
Total Alkalinity
Alkalinity test kit
Reprocessing area at each of the 11 point-of-use sites (3 - critical sinks, 1 - utility sink, 1 - RO storage tank, 4 - medivator sterilizers, 2 – washers). Immediately downstream of water treatment process.
Contractor
Monthly
FMS (contract)
Ionic contaminants
Specific tests for aluminum, chloride, copper, iron, manganese, Nitrate, Phosphate, Sulfate, Silicate, Zinc.
The main water feedline into the facility
Contractor
Annually
FMS (contract)
NOTE 1: The specifications presented in this table should be tested to the parameters specified in Table 1 and in accordance to ANSI/AAMI ST108: 2023 where applicable.
NOTE 2: The recommendations for frequency of testing in this table are the recommended minimum frequency. If problems or issues arise with the water quality, it might be necessary to increase the frequency until they are resolved.
Table 2: Overview of water quality monitoring, edited to include Responsible Party.
b. The successful bidder shall provide testing services for critical water (reverse osmosis) for bacteria (weekly), endotoxin (monthly), pH (meter) (monthly), total hardness (monthly), conductivity (monthly), total alkalinity (monthly), and ionic contaminants (annually) per Table 2 above.
c. Tests shall occur at the required locations per Table 2. The ‘main water feedline into the facility’ shall be tested at the domestic water supply inlet to the critical water system.
d. Water sampling shall be performed in accordance with ANSI/AAMI ST108:2023 Annex H.
4. Testing Days
a. Testing shall occur during normal business hours (8AM – 4:30PM).
b. Testing shall occur between the 10th and 15th day of the month.
c. The successful bidder shall notify the Contracting Officer’s Representative (COR) prior to travelling to the site to ensure that testing can be completed on the requested date. The COR may request to move the testing date due to mission requirements and this shall result in no additional cost to the government to reschedule.
5. Reporting
a. Reporting shall be per typical company reporting documentation. Additionally, documentation shall be reported per Attachment A (excel sheet).
b. Reports shall be provided through email to the COR before the last business day in the testing month.
c. Reports shall provide recommendations for remediation of deficient testing parameters.
d. Chain of custody docume…
Source: SAM.gov, as posted. Verify the current solicitation before responding.
Pursue this opportunity with Mindy
See who holds it now, who else is bidding, and draft your response — grounded in real government data, not generic AI.
View the original notice on SAM.gov ↗Similar Active Opportunities (NAICS 541330)
ADaPT Gen III PCB Fabrication/Assembly
NATIONAL AERONAUTICS AND SPACE ADMINISTRATION · Combined Synopsis/Solicitation · due July 30, 2026
NAVSUP Virtual Industry Day - NETSAFA International Training Center
DEPT OF DEFENSE · Sources Sought · due August 3, 2026
HAESI 2.0 INDUSTRY DAY 05 AUG 2026
DEPT OF DEFENSE · Special Notice · due August 3, 2026