NOAA: Calibration and Accuracy Temperature Bath for the NWS/Sterling Field Support Center (SFSC). in Sterling, VA
COMMERCE, DEPARTMENT OF
Notice type
Solicitation
Solicitation #
1305M226Q0240
NAICS
334516
PSC
6640
Set-aside
Small Business Set Aside - Total
Posted
August 7, 2026
Response due
August 19, 2026
Place of performance
Sterling, VA
What this opportunity is
The Department of Commerce is buying a calibration and accuracy temperature bath for the National Weather Service's Sterling Field Support Center in Virginia. This small business set-aside opportunity suits companies classified under NAICS 334516. As a solicitation notice, small businesses should track this opportunity to prepare a bid, given the specific equipment requirements outlined, including internal bath dimensions, temperature ranges, and stability criteria.
Analysis by Mindy, grounded in the SAM.gov notice.
Description
Statement of Work
142082-26-0415
SFSC Temperature Calibration Bath Replacement
The National Weather Service Sterling Field Support Center (SFSC) Temperature and Humidity
laboratory provides internal calibration for 50-60 platinum resistance thermometers and accuracy
comparison services for NWS programs testing new temperature sensors and equipment. In order
to provide accurate calibrations, the use of an extremely stable and accurate temperature source is
required. Staff utilize two stable temperature calibration bath sources, one for low temperature
ranges and another for high temperature ranges. Currently one of our temperature calibration baths
required for temperature accuracy testing and sensor calibrations is no longer working correctly.
The unit can no longer cool properly and based on in house assessments the unit will likely stop
functioning completely. The lack of system cooling also causes a major safety risk along with the
inability for the temperature bath to function in a correct manner. The unit must be replaced with
a new stable temperature source to continue to provide accurate calibration and temperature
accuracy testing for NWS programs and sensors.
Performance Criteria
The equipment requirements listed below must all be met for the final acceptance of the system.
Internal Bath Dimensions (Minimum): 6in x 6in x 8in
Internal Bath Dimensions (Maximum): 9in x 14.5in x 16in
Immersion Depths: 8in to 16in
Access Opening (Minimum): 6in x 6in
Access Opening (Maximum): 14in x 8.5in
Volume: 1 to 11 gallons
Bath Fluid Type: Must operate with Xiameter PMX-200 Silicone Fluid 1cSt at
temperatures between -80°C to 10°C and Distilled Water at temperatures between 5°C to
60°C
Low Temperature Range: Less than or equal to -80°C
High Temperature Range: Greater than or equal to 60°C
Power Source: 115 or 230 VAC
Environmental Operating Range:
o Relative Humidity: 80% at 31°C to 50% at 40°C
o Temperature : 5 to 50°C
Full Range Stability: Less than or equal to 0.02°C
Full Range Uniformity: Less than or equal to 0.02°C
Combined Stability and Uniformity: Less than or equal to 0.025°C
Heating Rate: Greater than or equal to 0.625°C / minute
-- 1 of 2 --
Cooling Rate: Greater than or equal to 0.4375°C / minute
Display Resolution: Less than or equal to 0.01°C
Set-Point Accuracy: Less than or equal to 0.5°C
Set-Point Repeatability: Less than or equal to 0.01°C
Serial communications:
o Must possess the ability to receive commands sent from a third-party terminal
program such as Hyper Terminal or PuTTy. These commands must have the
ability to control the operation of the temperature bath specifically the ability to
set and maintain a temperature setpoint. Must also have the ability to send
commands that can be received by a third-party terminal program such as Hyper
Terminal or PuTTy.
Overall Equipment Dimensions: All equipment must fit within 31in W x 31in L x 60in
H area
Maintenance: Maintenance costs minimized
Shipping: Shipping Cost included
Inspection and Acceptance
Upon completion, contractor shall notify the Government representative for approval and
acceptance. The acceptance period will include a full successful demonstration in accordance with
vendor supplied accepted test plan with the federal representative signature. Delivery of
acceptance test results from the test plan must be submitted with invoice for processing and
approval of payment.
Delivery/Point of Service to be Completed, if applicable:
Sterling Field Support Center
43741 Weather Service Rd
Sterling VA, 20166
National Weather Service POC:
Kathryn Franc
kathryn.franc@noaa.gov
301-388-8565
-- 2 of 2 --
Source: SAM.gov, as posted. Verify the current solicitation before responding.
Pursue this opportunity with Mindy
See who holds it now, who else is bidding, and draft your response — grounded in real government data, not generic AI.
View the original notice on SAM.gov ↗Similar Active Opportunities (NAICS 334516)
Gold Nanostructured Target
ENERGY, DEPARTMENT OF · Solicitation · due August 26, 2026
58--NOTICE OF INTENT TO AWARD A SOLE SOURCE
INTERIOR, DEPARTMENT OF THE · Special Notice · due August 28, 2026
UV-VIS Spectrophotometer and Autosampler
HEALTH AND HUMAN SERVICES, DEPARTMENT OF · Combined Synopsis/Solicitation · due September 4, 2026
Custom CryoAdvanced internals
DEPT OF DEFENSE · Special Notice · due August 24, 2026
Microbiology Analyzer and Reagents - Cost per Test
DEPT OF DEFENSE · Combined Synopsis/Solicitation · due August 31, 2026
20--eDNA AUTOSAMPLER
INTERIOR, DEPARTMENT OF THE · Special Notice · due August 26, 2026