What this opportunity is
The government is acquiring an Applied Biosystems QuantStudio 3 Real-Time PCR System for research purposes, specifically for the National Institute of Arthritis and Musculoskeletal and Skin Diseases (NIAMS) laboratories in Maryland. This opportunity is open to all businesses, as no set-aside is used. The notice type is a Combined Synopsis/Solicitation, indicating that interested parties should track the solicitation for updates rather than submitting a bid immediately.
Analysis by Mindy, grounded in the SAM.gov notice.
Description
Requirements:
To ensure continuity of research, NIAMS proposes the acquisition of an Applied Biosystems QuantStudio 3 Real-Time PCR System. qPCR is a core technology used for nucleic acid amplification, quantification, and analysis and supports a wide range of research applications, including gene expression studies, pathogen detection, transcript quantification, and validation of experimental results. The QuantStudio 3 is the closest replacement to the current StepOnePlus systems, allowing researchers to maintain established workflows and consistency in experimental design and data interpretation. The QuantStudio 3 offers several operational advantages, including user-friendly software, compatibility with both Mac and PC platforms, touchscreen operation, and the ability to function as a standalone instrument or in conjunction with a computer. Its thermal block design is similar to the existing systems, minimizing the learning curve for users and supporting a broad range of genomic applications. In addition, the manufacturer offers trade-in credit for the current instruments, reducing the overall acquisition cost while providing access to a fully supported platform with available parts and technical support. Acquiring the QuantStudio 3 at this time will help prevent interruptions to research activities, maintain reliable and consistent data generation, and ensure that NIAMS laboratories have access to a supported and dependable qPCR platform.
4.1 PRODUCT FEATURES/SALIENT CHARACTERISTICS The contractor shall provide one Applied Biosystems QuantStudio 3 Real-Time PCR System or an equal system that meets all the following minimum requirements. 4.1.1. System Features/Benefits The system shall: • Utilize a platform providing 1 TB or storage space for data sharing at user discretion • be used with online scientific apps • come with desktop-based software, allowing single-plate run setup and analysis.
4.1.2. Technical specifications – Instrument The instrument shall: • have a 96 well 0.1 ml plate block format and the option to run fast protocols <30 minutes • not require a field service engineer for set up by end user. Set up should be easy for end user – directly out of the box. • not require a computer for installation • use an RNase P verification plate to verify performance • not require dye calibration upon installation • have dimensions fo 27 x 50 x 40 cm and be designed to maximize horizonal and vertical bench space • be capable of analyzing 96 samples simultaneously in various formats incluing both 96-well plates, 8-tube strips and individual tubes • the 96-well block should be made of at least 3 separated Peltier elements – VeriFlex Blocks. The maximum temperature difference programmable across the block should be 10 degrees C. 5 degrees C should be the maximum difference in temperature allowed between adjacent blocks. • Support reaction volumes of 10-30 uL on 0.1mL the 96-well block • Be factory calibrated with FAM, SYBR Green 1, VIC, NED, ABY, JUN, TAMRA, and ROX dyes • Have melt or dissociation curves that take place in steps with deltaT varying between 0.015 degrees C and 3.66 degrees C • Have a maximum block ramp rate of 9.0 degrees C/sec for the 0.1 mL block • Have a block with a temperature range of 4-99.9 degrees celcius • Have a heated cover with temperature range of 50-110 degreesC 2
• Have a long-lasting bright white LED source with median lifetime of at least 60,000 hours • Have a CMOS imager for data acquisition and collects data for each filter combination in less than 2 seconds • be used with 120-240 V outlets without modification or special plug requirements • Include a heated lid assembly to heat tops of sample plates and minimize reaction evaporation.
4.1.3 Technical specifications – Instrument onboard computer The system shall: • Possess the capability to run in stand-alone mode, without an attached computer • Include a touchscreen interface and store protocols for quick starts with no computer needed • Have a touchscreen interface that enables viewing amplification plots and filtering them by sample, target and task • Have a touchscreen interface that facilitates uploading of protocols or downloading of run files via a USB, local network and the Connect platform • Have a touchscreen interface that allows uploading of protocols and run files directly to the Connect platform • Contain onboard storage with memory capacity of 10 GB for storing 1,000-5,000 standard absolute quantification run files, and be designed for run protection in case of power outages • Be able to be customized through the touchscreen for time zones and service/calibration notification settings • Have a touchscreen interface that supports the creation of local user accounts with PIN-protected access that can be linked to the Connect platform to online accounts on thermofisher.com • Have a touchscreen interface that allows creation and running of protocols as guests • Have a touchscreen interface that allows for editing of protocols at all stages of the reaction cycle as well as block settings, temperature settings duration, ramp rate, etc. • Have a touchscreen interface that allows a template to be loaded from a USB, local network or the Connect platform • Have a touchscreen interface that allows user to enter reagent information • Have a touchscreen interface that allows a run to be paused and during that pause for the block temperature to be set as well as for the drawer to be opened and closed • Have a touchscreen interface that allows editing of sample and target assignments • Have a touchscreen interface that allows the repeat of a previous run 3 • Have a touchscreen interface that allows users to perform dye calibrations and view results on the touchscreen • Have a touchs…