Deionization System with Recirculation and Monitoring Skid 2
ENERGY, DEPARTMENT OF
Notice type
Solicitation
Solicitation #
485281
NAICS
221310
PSC
4610
Posted
July 24, 2026
Response due
July 31, 2026
Place of performance
Oak Ridge, TN
What this opportunity is
The Department of Energy is buying a deionization system with recirculation and monitoring skid, which suits businesses classified under NAICS code 221310. This solicitation is for a specific type of equipment, including analytical transmitters for comprehensive liquid analysis. As a solicitation notice, businesses should track this opportunity to prepare for a potential bid.
Analysis by Mindy, grounded in the SAM.gov notice.
Description
Analytical Transmitters
M300 Transmitters
for Comprehensive Liquid Analysis
THORNTON
Leading Pure Water Analytics
M300 Transmitters
Conductivity / Resistivity
pH / ORP
Dissolved Oxygen
Dissolved Ozone
Flow / Total Flow
-- 1 of 12 --
M300 Transmitters
2
The METTLER TOLEDO THORNTON M300 series of process analytical instruments provide
single-channel, multi-channel and multi-parameter models to measure conductivity/ resis-
tivity, pH/ORP, dissolved oxygen, ozone and flowrate. Both ¼ DIN and ½ DIN enclosures
are available for especially convenient panel, wall and pipe mounting.
The standard for measurement
in pure water treatment
Parameter Channels Outputs Specs on page*
Conductivity / Resistivity 1 2 analog; 4 relays 3
Conductivity / Resistivity 2 4 analog; 4 relays 3
pH ORP 1 2 analog; 4 relays 4
Conductivity / Resistivity, pH / ORP
Dissolved Oxygen, Ozone 2 4 analog; 6 relays 5 - 6
Flow 1 2 analog; 4 relays 4
Flow 4 4 analog; 4 relays 4
*physical and electrical specifications on page 7, order numbers on page 10
Model selection
Features
Large backlit 4-line display
Compact ¼ DIN panel-mounting with NEMA 4X, IP65 panel seal
Robust ½ DIN wall-mounting with NEMA 4X, IP65 back cover
User-selectable dual-level password protection
On-line sensor diagnostics for pH
Internal PID control with relay or analog output
USB port for configuration and data acquisition
Direct sensor inputs for all parameters
Compatible with existing Thornton sensors
Plug-in terminal connectors for easy wiring
Universal AC/DC power supply
Multiple languages: English, French, German, Italian, Spanish
CE compliant, UL Listing pending for US and Canadian standards
Applications
Pure and ultrapure water treatment for semiconductor rinsing, critical power/steam makeup and pharmaceutical waters.
Semiconductor processing in rinsers and wet benches with precise resistivity measurement and temperature
compensation.
Power plant cycle chemistry and stator cooling monitoring with exceptionally accurate temperature compensation for
specific and cation conductivity and pH and very low maintenance dissolved oxygen measurement.
Pharmaceutical water monitoring to meet USP, EP, and JP conductivity requirements including built in tables of
alarm limits.
Reclaim, recycle and wastewater treatment for the above industry applications for contaminant detection,
diversion and neutralization.
-- 2 of 12 --
3
One & Two channel models
Highest accuracy measurement & temperature compensation
Wide range of measurement allows verification in ASTM
standards even for pure water measurements
Two- & four-electrode conductivity/resistivity measurements
Reverse osmosis % rejection computation on two-channel models
Direct acid & caustic concentration readout
USP ‹645› & EP alarm setpoints
Measurement Specifications
M300 Conductivity / Resistivity Models
Functional
Conductivity/Resistivity Ranges
0.01 Constant Sensor 0.002 to 200 μS/cm (5000 ohm-cm to 500 Mohm-cm)
0.1 Constant Sensor 0.02 to 2000 μS/cm (500 ohm-cm to 50 Mohm-cm)
10 Constant Sensor 50 to 40,000 μS/cm (25 to 20 Kohm-cm)
4-electrode Sensor 0.01 mS/cm to 650 mS/cm (1.54 ohm-cm to 100 Kohm-cm)
readout in equivalent S/m ranges is selectable
Concentration Ranges of HCl, NaOH, H2SO4 0-20%, 0-15%, 0-20%
TDS Ranges (CaCO3 and NaCl) cover equivalent conductivity ranges
Calculated Paraameters (2-channel) % Rejection, power plant calculations of pH based on specific and cation conductivity, and C02 based on
cation and degassed conductivity
Resolution four significant digits, auto-ranged
Update rate display and outputs, once per second
Temperature Measurement Range -40 to 200°C (-40 to 392 °F), resolution 0.1°
Temperature Sensor Input RTD, Pt1000 (Pt100 with adapter)
Temperature Compensation selectable as: Std (standard high purity Thornton/Light), Light 84, Std referenced to 75°C,
linear %/°C, 50% glycol, 100% glycol, cation, ammonia, isopropyl alcohol, none
Sensor maximum distance 61 m (200 feet); 15 m (50 ft) with 4-electrode sensors
Performance
Conductivity/Resistivity Accuracy ± 0.5% of reading or 0.5 ohm, whichever is greater, up to 18 Mohm-cm
Repeatability ± 0.1% of reading
Temperature Relative Accuracy ± 0.25 °C (± 0.45 °F)
Temperature Resolution 0.01 °
Temperature Repeatability ± 0.13 °C (± 0.23 °F)
Outputs – see p.7 for ratings Single channel Two channel
Setpoints/Alarms 2 - high, low, outside, between, USP, EP 6 - high, low, outside, between, USP or EP
Relays 2 SPDT, 2 SPST reed 2 SPDT, 1 SPST NO, 1 SPST NC, 2 SPST reed
Analog Output Signals 2 4
Discrete Inputs 1 2
-- 3 of 12 --
M300 Transmitters
4
M300 pH/ORP Models
Functional
pH, ORP Ranges -1.00 to 15.00 pH, -1500 to 1500 mV
Temperature Range - 30 to + 130°C (-22 to 266 °F)
Temperature Sensor RTD, Pt1000 (Pt100 with adapter)
Temperature Compensation automatic / manual for electrode output, plus adjustable solution
temperature coefficient for solution ionization effects
Update rate display and outputs, once per second
Calibration 1- or 2- point, with auto buffer recognition
Diagnostics selectable continuous checking of membrane resistance and reference
diaphragm/junction resistance (with solution ground sensors)
Sensor Maximum Distance 10 m (33 ft)
Performance
pH, ORP Relative Accuracy ± 0.03 pH, ± 2 mV
pH, ORP Resolution 0.01 pH, 1 mV
Temperature Accuracy ± 0.25 °C (± 0.45 °F)
Temperature Resolution 0.1 °C
Outputs – see p.7 for ratings
Setpoints/Alarms 4 - high, low, outside, or between
Relays 2 SPDT, 2 SPST reed
Analog Output Signals 2
Discrete Inputs 1
Direct electrode input
On-line sensor diagnostics for measuring and
reference electrode
Auto buffer recognition with extensive buffer library
Internal PID control
Combined high/low setpoint for alarming outside a band
Solution temperature compensation for pure water
ionization effects as well as conventional electrode
temperature compensation
Automatic sensor cleaning
Measu…
Source: SAM.gov, as posted. Verify the current solicitation before responding.
Pursue this opportunity with Mindy
See who holds it now, who else is bidding, and draft your response — grounded in real government data, not generic AI.
View the original notice on SAM.gov ↗Similar Active Opportunities (NAICS 221310)
IRRIGATION SYSTEM FOR NATL WEATHER SVC IN TEXAS
COMMERCE, DEPARTMENT OF · Combined Synopsis/Solicitation · due August 7, 2026
Class B Water Testing Solicitation
JUSTICE, DEPARTMENT OF · Solicitation · due December 19, 2025
DINO: WATER RESEVOIR INSPECTIONS
INTERIOR, DEPARTMENT OF THE · Solicitation · due August 11, 2026
J047--Water Piping Rebuild
VETERANS AFFAIRS, DEPARTMENT OF · Combined Synopsis/Solicitation · due August 7, 2026
Z--DIAGNOSE AND REPAIR PILLAR WELL, Pinnacles NP
INTERIOR, DEPARTMENT OF THE · Award Notice
J046--Installation and Maintenance for Water Monitoring System at the Bedford VA Medical Center
VETERANS AFFAIRS, DEPARTMENT OF · Combined Synopsis/Solicitation · due August 13, 2026