Gas Chromatograph Single Quadruple Mass Spectrometer
AGRICULTURE, DEPARTMENT OF
Notice type
Combined Synopsis/Solicitation
Solicitation #
1232SA26Q0925
NAICS
334516
PSC
6640
Posted
July 27, 2026
Response due
August 7, 2026
Place of performance
Fargo, ND
Description
SPECIFICATIONS
Specifications Description:
The USDA ARS Biosciences Research Laboratory, Food Animal Metabolism Research Unit and the Sugarbeet Research Unit, Fargo, North Dakota have need for a bench-top single quadrupole mass spectrometer configured with a gas chromatograph with autosampler for liquid injection, pyrolyzer unit with components, and solid phase micro-extraction capabilities. A satisfactory bid must meet the following requirements:
Single Quadrupole Mass Spectrometer with following; Quantity 1, Unit Each:
Inert electron ionization (EI) source
Electron multiplier detector
Ion gauge
Foreline vacuum pump - dry scroll pump or equivalent
Turbo High Vacuum Pump
Maximum column ID of 0.32 mm or greater
Setup for Helium Carrier gas
Acquisition: Full ion scan and selected ion monitoring (SIM) capability
Mass Range: At least 1.2 – 1000 atomic mass units or higher high mass limit
Scan speed of 20,000 u/second (u=unified atomic mass units)
All accessories and kits needed for install and operation
Uninterruptable Power supply, 6kV; Quantity 1, Unit Each
Gas Chromatograph; Quantity 1, Unit Each:
Capillary flow technology
GC Diagnostics functionalities
2 injector ports, specified below
Maximum temperature ramps of at least 90°C/min
Operating temperature range
At least 3-4°C above ambient temperature to 450°C
Oven Cool down
450°C to 50°C in 4.0 minutes (22°C ambient)
Peak Area Repeatability <0.5% RSD
Retention Time Repeatability <0.01%
Typical Pressure Control <0.01 psi
Oven exhaust deflector
All accessories and kits needed for install and operation
Autosampler and Injector Ports; Quantity 1 with components, Unit Each:
Standard liquid and headspace (gas) autosampler injector and vial trays (2mL and 10mL) with smart rail system for multipurpose use components below; quantity 1, unit each
Universal sample tray holder with trays for 2 mL vials and 10 mL vials
Universal syringe module for liquid injection 10 microliter or less
Standard syringe wash station
Sample vial tray Pyrolysis injection (see pyrolyzer specifications)
Split/Splitless injector modules; quantity 2, unit each
Install kits for 2 injector ports; quantity 2, unit each
Adaptor kit for gas injection for pyrolyzer on split/splitless injector; quantity 1, unit each
Solid phase micro extractor system (SPME) ; Quantity 1 with components below, Unit Each:
SPME fiber conditioning station
SPME fiber kit with fibers and test mixture for autosampler and GC
SPME heater/agitator for 10 mL vials
Adapters for heater/agitator
Pyrolyzer Unit and Components; Quantity 1 with components, Unit Each:
Multi Shot Pyrolyzer with install kit; quantity 1, unit each:
Must consist of a ceramic micro furnace design with temperature reproducibility of +/- 0.1ᵒ C or equivalent.
Thermocouple temperature measurement is designed into the micro furnace or equivalent for temperature calibration.
Unit mounts onto the GC inlet system with no external transfer lines.
Settable and programmable from ambient room temperature + 10ᵒ C up to 1050ᵒ C in 1ᵒ C increments with +/- 0.1ᵒ C stability.
GC Interface must be settable from 40ᵒ C to 450ᵒ C in 1ᵒ C increments with +/- 0.1ᵒ C stability.
Must meet the following guaranteed performance requirements:
Using flash pyrolysis of 25 µg of a polystyrene solution containing methyl stearate (5 µg/µl) at 550ᵒC, the relative standard deviation (RSD%) of the peak area ratios of the styrene trimer and methyl stearate must be ≤ 2%.
The cooling time of the furnace or equivalent from 800ᵒC to 50ᵒC must be ≤ 10 minutes to maximize sample throughput.
Must include a 2-year warranty.
Needs for Auto Sampler
A single sample tray must be capable of holding 48 samples; quantity 1, unit each
Sample cups must be made from stainless steel that has been deactivated.
Sample cup volume must be ≤ 80µl.
Must allow automated unattended analysis of 48 samples, including single-shot analysis and double-shot analysis.
Cryogenic Trapping (Cryo-Trap) ; quantity 1, unit each:
A Cryo-Trap device utilizing liquid nitrogen (LN2).
The Cryo-Trap must be able to reach temperatures as low as -196ᵒC so that CO2 may be effectively cryo-trapped.
The Cryo-Trap must spray LN2 directly onto the analytical column in a very narrow band of < 5mm.
The Cryo-Trap must utilize gaseous nitrogen flow through a thermal exchange coil (placed in LN2) with the ability to vary the flow rate from 3-7 L/min, allowing choice of desired cooling temperatures.
A optional thermal exchange coil must be available for use with a 30L LN2 Dewar (user-supplied) to allow unattended operation with the Auto Sampler.
Cryo-Trap must be software-controlled using the same software as the pyrolyzer.
Multi-Functional Splitless sampler (see above); quantity 1, unit each:
Used with the Multi-Shot Pyrolyzer, which allows for the F-Splitless injection method, backflush, solvent cut, and improves peak separation.
This device must allow the user to perform splitless analysis for Flash pyrolysis (Single-shot analysis) and Double-shot analysis
Software; quantity 1, unit each (see below in workstation section):
Full license
Must be capable of controlling all set points of the pyrolyzer and the autosampler or work seamlessly with GC-MSD software.
Must be capable of controlling the multiple modes of operation, including:
Evolved Gas Analysis (EGA)
EGA is defined as analyzing a sample using a direct connection from the S/SL GC injection port and the MS using a 2.5 meter, 0.15mm i.d. deactivated stainless steel capillary tube.
Pyrolysis (PY)
Reactive Pyrolysis (RxPY)
Heart cuts based on the EGA Thermograms
High pressure, high temperature reaction chemistry
Search Library; quantity 1, unit each:
Software for microplastic library searching must be available with the following specifications:
Incorporate search algorithms that allow spectral library searching of pyrograms based on C1-C10, C1-C20, and C1-C40 carbon groupings,
Automaticall…
Source: SAM.gov, as posted. Verify the current solicitation before responding.
Pursue this opportunity with Mindy
See who holds it now, who else is bidding, and draft your response — grounded in real government data, not generic AI.
View the original notice on SAM.gov ↗Similar Active Opportunities (NAICS 334516)
iCAP-TQ additional gas module (Toxic Elements)
HEALTH AND HUMAN SERVICES, DEPARTMENT OF · Solicitation · due July 31, 2026
6515--Handheld narcotics analyzer
VETERANS AFFAIRS, DEPARTMENT OF · Combined Synopsis/Solicitation · due August 3, 2026
Purchase of UHV Components for ICEE facility
NATIONAL AERONAUTICS AND SPACE ADMINISTRATION · Combined Synopsis/Solicitation · due July 29, 2026
Image Intensifier Tube with Housing
ENERGY, DEPARTMENT OF · Solicitation · due August 10, 2026
annual maintenance and service agreement for autoclave
NATIONAL AERONAUTICS AND SPACE ADMINISTRATION · Special Notice · due July 29, 1016
Procure a new WD-XRF for Color Certification Analysis.
HEALTH AND HUMAN SERVICES, DEPARTMENT OF · Combined Synopsis/Solicitation · due August 7, 2026