The NATIONAL AERONAUTICS AND SPACE ADMINISTRATION is procuring a NETZSCH Simultaneous Thermal Analyzer, specifically the STA 509 Jupiter model, for the Johnson Space Center Astromaterials Research & Exploration Sciences Division. This 8(a) Sole Source procurement is suitable for small businesses certified under the 8(a) Business Development program, which is a set-aside program for historically underutilized businesses. The notice type is a Solicitation, indicating that interested parties should track the opportunity for updates rather than preparing a bid.
Analysis by Mindy, grounded in the SAM.gov notice.
Description
STATEMENT OF WORK
1. Objective/Requirements
The NASA Johnson Space Center Astromaterials Research & Exploration Sciences
Division requires the procurement of a NETZSCH Simultaneous Thermal Analyzer.
2. Characteristics, Scope, and Specs
LINE PART NO. DESCRIPTION QTY.
1 STA509SEA13.000-
00
STA 509 Jupiter® Simultaneous Thermal Analyzer
(TG-DSC/DTA Apparatus) Temperature range
according to selected furnace consisting of:
STA509SEA13.000-00
STA 509 Jupiter Select ASC automatic sample
changer, "Heavy" Basic system with top-loading
micro balance, electrical temperature control unit
for the balance chamber, electric motor driven
furnace hoist and built-in furnace control and
power supply, programmable gas flows for 3 or 4
gases. The system is suitable for measurements in
normal atmosphere, in dynamic and static inert gas
atmosphere. The instrument is vacuum-tight (10E-4
mbar, achievable vacuum depending on the actual
evacuation system). Equipped with:
- Automatic furnace hoist "Heavy" prepared for
supporting a heavy maximum temperature furnace
and the automatic sample changer
- Automatic sample changer (ASC) for automatic
handling of up to 20 crucibles, with removable
magazine, including 20 alumina sample crucibles
GB399972 (85μl), complete with control
electronics (expandable with a separately available
piercing add-on)
- Colored touch display for important functions for
measurement and calibration
- Colored LED status bar with clear indication of
the instrument status and measurement progress
- Integrated TA system controller for data
acquisition, temperature control, gas flow control
- Temperature range: -150°C...2000°C (depending
on the selected furnace)
- Heating rates: 0.001...1000 K/min (depending on
furnace selected)
- Balance system Weighing range: 35 g Digital
resolution: 0.1 μg Initial sample weight: 35 g max.
(incl. sample pan)
1
2 SW-9-STA509SE SW-9-STA509SE 1
-- 1 of 11 --
Proteus basic software for STA 509 Select for data
acquisition, storage and evaluation under MS
WINDOWS with ExpertMode user interface.
ExpertMode allows access to the complete
software functionality, many settings and detailed
help documentation. Contains AutoEvaluation (for
TGA and DSC) for automatic evaluation of effects,
BeFlat® (for TGA and DSC) for automatic
buoyancy and baseline correction, as well as the
Report Generator for simple and customizable
creation of reports. Detailed information: see
technical data sheet STA software Proetus 9.x
(Hardware requirements: up-to-date PC with MS
Windows per technical datasheet and printer)
3 SW-EVDSC-
OPTION
Proteus option AutoEvaluation for DSC measuring
data (DSC AutoEvaluation) The Software
extension "DSC AutoEvaluation" finds the thermal
effects on the measured DSC data for heating
segments with stable baseline before the first
thermal effect and after the last thermal effect.
Suitable for automatic evaluation of metal melting,
which is used for calibration procedure, for
evaluation of polymers and in general for endo and
exothermic effects upon heating.
Furthermore for the determination of
the oxidation induction time or temperature (OIT)
based on tangential
and offset methods. The AutoEvaluation function
"UV-Curing" significantly simplifies the evaluation
of measurements of light-curing
materials: correction of the heat introduced by the
light pulse is automatically performed, as well as
an evaluation of the partial enthalpies, a display of
the total reaction enthalpy, and the percentage of
cross-linking.
For melting of metals the evaluation range, onset
and peak area for one
melting peak are found. For polymers, the effects
such as melting, glass
transition, crystallization, curing, evaporation of
additives, melting of plasticizers, multiple melting
peak and multiple glass transition are
found. For each thermal effect its type is defined,
the evaluation range is found automatically, and
automatic evaluation is done corresponding to
detected effect type. Automatic evaluation of endo-
and exothermic effects for STA instruments
can be set as dependent on the mass changes
evaluated.
-- 2 of 11 --
In general, one can select which effects are
evaluated and which effect
properties are shown; furthermore one can change
the thresholds of the
sizes of the effects to be evaluated. Results of
AutoEvaluation are independent from operator, and
serve as the verification for experts.
After performing of automatic evaluation its results
can be edited manually.
(Required: Most recent version of Proteus basic
software,version 9.x and up)
4 SW-EVTGA-
OPTION
Software extension TGA Auto-Evaluation for
Proteus 9 The Software extension "TGA
AutoEvaluation" finds the individual mass change
steps on the measured TGA data for heating,
cooling, isothermal segments and for complete
measurement curve of several different segments.
Suitable for automatic evaluation of mass change
including both mass loss and mass gain. TGA
Autoevaluation contains analysis of TGA curve
and automatic calculation of DTG curve and
analysis of it. For each mass change step the
evaluation range is found automatically, and
automatic evaluation of it is done on TGA and
DTG curves. Overlapping steps are evaluated
separately if they have a well-visible peak or a well
visible shoulder on DTG data. One can change the
threshold of the size of mass changes to be
evaluated. Furthermore, one can select if the DTG
curve and the residual mass are evaluated and
shown. Results of TGA AutoEvaluation are
independent from operator, and serve as the
verification for experts. After performing of
automatic evaluation, its results can be editable
manually. (Required: Proteus basic software
version 9.x)
1
5 SW-DSCBFLT-
OPTION
Software extension "DSC-BeFlat" for Proteus
Software extension "DSC-BeFlat" for correction of
temperature and heating rate dependent DSC
baseline deviations applying a physical model to
achieve highest baseline stability with minimal
curvature in wide temperature ranges. The
correction is alread…
Source: SAM.gov, as posted. Verify the current solicitation before responding.