Commercial trace gas analyzer
COMMERCE, DEPARTMENT OF
Notice type
Combined Synopsis/Solicitation
Solicitation #
1333ND26QNB030493U
NAICS
334516
PSC
6640
Set-aside
No Set aside used
Posted
August 26, 2026
Response due
August 27, 2026
Place of performance
Gaithersburg, MD
Description
TEMPLATE
STATEMENT OF WORK
Title: Commercial Trace Gas Analyzer
Lab: Chemical Sciences Division
I. BACKGROUND INFORMATION
The semiconductor supply chain is global, specialized, and interconnected. Chipmakers
do business with thousands of individual suppliers that provide the highly complex
materials and tools used to produce semiconductors. To address the lack of full visibility
into the semiconductors markets supply chain and R&D ecosystem gaps NIST will
conduct the measurement science, or metrology, critical to the development of new
materials, packaging, and production methods in chip manufacturing.
As part of CHPS Grand Challenge 1: Metrology for Materials Purity, Properties, and
Provenance, we will create measurement technology, reference data and materials to
perform trace impurity gas analysis at new levels of precision. Success will enable
next-generation diagnostics to reliably identify impurities in process gases at sub-parts-
per-billion (<1 ppb) levels. Results will drive partners to develop industry ready
reference materials, point-of-use diagnostics and models to achieve breakthroughs in
critical tooling like extreme ultra-violet lithography. Our diagnostics and metrology-grade
data will enhance quality systems by establishing a framework for impurity fingerprinting
with traceability to new gas primary standards, reference materials and documentary
standards.
NIST requires a commercial trace gas analyzer to serve as a benchmark for newly
developed ultra-sensitive optical diagnostics. Current optical diagnostics relied upon by
the semiconductor industry suffer from trade-offs between sensitivity and quantitation.
This analyzer will be used in tandem with NIST-developed cavity ring-down
spectroscopy (CRDS) to verify the performance of new analytical instrumentation and
contribute to the certification of gas reference materials (RMs). This acquisition supports
the CHIPS R&D Metrology Program's goal of establishing reliable data for trace
impurities.
II. PURPOSE
The Chemical Sciences Division (CSD) is part of the Materials Measurement Laboratory
(MML) at NIST. MML serves as the national reference laboratory for measurements in
the chemical, biological, and materials sciences. MML provides the measurement
science, standards, technology, and data required to support the nation's need to
design, develop, manufacture, and use materials. This acquisition directly supports the
Division's mission by providing a standardized tool for the verification of primary gas
standards and the characterization of new high-quality semiconductor and electronics
gases.
The Gas Sensing Metrology Group (GSMG) of the CSD will develop primary standards
and reference materials for “zero gas” semiconductor and electronics-grade gases.
-- 1 of 8 --
2 | P a g e
Development of such standards will require specialized instrumentation to properly
separate, identify, and quantify trace impurities at sub-parts-per-billion levels. One of
the instruments needed is a trace gas analyzer. This system will be utilized to confirm
impurities in pure semiconductor grade gases purchased from vendors to qualify for use
in gas mixture preparations, to verify the performance of new analytical instrumentation
being developed, to validate new primary gas standards blended in-house, and to
contribute to the certification of gas reference materials.
III. MINIMUM REQUIREMENTS
The Contractor shall provide a system that meets all technical specifications identified
below. All items must be new. Used or remanufactured equipment will not be
considered for award. Experimental, prototype, or custom items will not be considered.
The use of “gray market” components are not acceptable. All line items shall be
shipped in the original manufacturer’s packaging and include all original documentation
and software, when applicable.
Line Item 0001:
Description: Trace Gas Analyzer
Quantity: One (1)
A. Technical Specifications
a. System
i. The trace gas analyzer shall be configured for the analysis of gases
to be delivered through a regulator from a gas cylinder.
ii. The trace gas analyzer shall also have capability for online
sampling, with gas flow delivered by a mass flow controller.
iii. The trace gas analyzer shall have a detection limit of less than 1
part-per-billion (< 1 ppb) of impurities including, but not limited to,
hydrogen (H2), carbon monoxide (CO), Nitrogen (N2), carbon
dioxide (CO2), and argon (Ar) in pure gases including, but not
limited to, helium (He), N2, and Ar.
iv. The trace gas analyzer shall provide automated data acquisition
suitable for routine laboratory diagnostics.
b. Software
i. The trace gas analyzer shall have software to control all functions.
ii. Software documentation shall provide the user with information on
how to control sample sequence, and method development.
Line Item 0002
Description: Operations and Maintenance Manual
Quantity: One (1)
a. Manual
iii. The contractor shall provide paper and/or electronic versions of the
operations and maintenance manuals, which cover proper
operation, routine maintenance, and troubleshooting for all system
components and controlling software.
-- 2 of 8 --
3 | P a g e
IV. SCHEDULE OF DELIVERABLES
Define all deliverables and indicate (1) the quantity desired; (2) the format in which each
deliverable is to be supplied; (3) the delivery date (based on the number of
weeks/days); (4) the place of delivery. The context of each deliverable shall be clear.
Keep in mind that every task does not require a deliverable. Deliverables are
always tangible items (i.e., report, minutes, etc.)
Complete the table providing the information for all the tasks identified above.
Deliverable
Number Description Quantity Due Date Place of Delivery
1 Trace Gas
Analyzer
One (1) No later than
90 days from
the date of
award.
NIST (Gaithersburg)
2 Operations and
maintenance
manual
One (1) No later than
90 days from
the date of
award.
NIST (Gaithersburg) or
electronic delivery if
…
Source: SAM.gov, as posted. Verify the current solicitation before responding.
Pursue this opportunity with Mindy
See who holds it now, who else is bidding, and draft your response — grounded in real government data, not generic AI.
View the original notice on SAM.gov ↗Similar Active Opportunities (NAICS 334516)
NSWCCD TA Instruments Discovery DMA 850
DEPT OF DEFENSE · Combined Synopsis/Solicitation · due September 1, 2026
Notice of Intent to Sole Source: Acquisition of Gel/Membrane Digital Imager and Analysis System
AGRICULTURE, DEPARTMENT OF · Special Notice · due September 1, 2026
Sodium Iodide Detector for Thyroid Counting
HEALTH AND HUMAN SERVICES, DEPARTMENT OF · Combined Synopsis/Solicitation · due August 31, 2026
sCMOS Scientific Camera, No Equivalents or Substitutions
ENERGY, DEPARTMENT OF · Combined Synopsis/Solicitation · due September 1, 2026