What this opportunity is
The Department of Health and Human Services is seeking a contractor for full-service maintenance, inspection, calibration, and repair of a body plethysmograph system at the Lawton Indian Hospital in Oklahoma. This opportunity is set aside for small businesses, making it suitable for firms specializing in medical equipment maintenance under NAICS 811210. Interested vendors should note that this is a Combined Synopsis/Solicitation, indicating they should track the notice for potential bidding opportunities rather than submitting a proposal at this stage.
Analysis by Mindy, grounded in the SAM.gov notice.
Description
Department of Health and Human Services
Indian Health Service
Performance Work Statement (PWS)
Lawton Service Unit – Respiratory Therapy MGC Diagnostic Platinum RTD
Elite Series Service Contract 11/03/2026
1. PURPOSE / BACKGROUND
The Indian Health Service (IHS), an operating division within the U.S. Department of Health and
Human Services (HHS), is responsible for providing federal health services to American Indians
and Alaska Natives. The mission of IHS is to raise the physical, mental, social, and spiritual
health of American Indians and Alaska Natives to the highest level and to assure that
comprehensive, culturally acceptable personal and public health services are available and
accessible to members of the 574 federally recognized Tribes across the United States.
The Lawton Service Unit (LSU) is a Joint Commission–accredited federal healthcare facility in
Oklahoma. The mission of the Lawton Service Unit is to deliver quality care in order to attain
positive health outcomes for all valued patients with emphasis on cultural sensitivity and human
dignity. To support this mission, the Respiratory Therapy Department utilizes specialized
pulmonary diagnostic equipment to assess and treat cardiopulmonary conditions.
The purpose of this Performance Work Statement (PWS) is to establish a full-service
maintenance, inspection, calibration, and repair agreement for the Respiratory Therapy
Department’s MGC Diagnostics Platinum Elite DL Series w/ Body Plethysmograph system
located at the USPHS Lawton Indian Hospital. This PWS defines the required outcomes; except
where specific technical or manufacturer standards are mandated, the methods used to achieve
those outcomes are the responsibility of the Contractor.
2. SCOPE / REQUIRED OUTCOMES
The Contractor shall provide all necessary labor, certified technicians, original equipment
manufacturer (OEM) replacement parts, calibration standards, testing equipment, travel, and
shipping necessary to maintain the MGC Diagnostics Platinum RTD Elite Series with Body
Plethysmograph system in optimal, certified operating condition.
All preventative maintenance, calibrations, diagnostic tests, and system recertifications shall
strictly adhere to Medical Graphics Corporation (MGC) manufacturer specifications, applicable
Federal regulations, and The Joint Commission accreditation standards.
Location of Service:
USPHS Lawton Indian Hospital
Respiratory Therapy Department
1515 Lawrie Tatum Road
Lawton, Oklahoma 73507
-- 1 of 11 --
Department of Health and Human Services
Indian Health Service
3. PERFORMANCE REQUIREMENTS AND STANDARDS
The Contractor’s performance will be evaluated against the following standards. (See Section
10 for the Quality Assurance / Surveillance Matrix.)
# Required Service
(Outcome)
Performance
Standard
Acceptable
Quality Level
(AQL)
Method of
Surveillance
3.1
Annual
Scheduled
Preventative
Maintenance
(PM) &
Recertification
Complete annual
factory-level PM,
comprehensive multi-
point inspection,
calibration, and full
system recertification
per MGC
specifications.
Performed within
±30 calendar
days of the
scheduled annual
anniversary date;
100% compliant
with OEM criteria.
Detailed Field
Service Report
& Biomedical
verification.
3.2
Unscheduled
Emergency &
Corrective
Repairs
On-site response and
diagnostic correction
for unexpected
system malfunctions
or hardware failures
using OEM
components.
On-site technician
arrival within 24 to
48 hours from
initial LSU service
call; system
restored to
operational
readiness.
LSU Bio-Med
service ticket
logs &
Contractor work
orders.
-- 2 of 11 --
Department of Health and Human Services
Indian Health Service
# Required Service
(Outcome)
Performance
Standard
Acceptable
Quality Level
(AQL)
Method of
Surveillance
3.3
Replacement
Parts & Hardware
Warranty
All replacement
components, circuit
boards, and sensors
must be new OEM
parts manufactured or
authorized by Medical
Graphics Corporation
(MGC).
100% OEM parts;
minimum 1-year
(12-month)
warranty on all
installed parts and
associated labor.
Work order
itemization,
parts receipts,
and invoice
review.
3.4
Technical Phone
& Remote
Diagnostics
Support
Direct access to
certified technical
specialists for real-
time troubleshooting,
system error
resolution, and
operational guidance.
Priority toll-free
phone and remote
assistance
available 6:00 AM
– 8:00 PM CST,
Monday through
Friday (excluding
federal holidays).
Call response 1
hour.
LSU staff
trouble logs and
vendor
teleconference
records.
3.5
Service
Documentation
& Joint
Commission
Compliance
Comprehensive,
itemized Field Service
Reports detailing
inspection items, as-
found/as-left test
results, calibrations,
replaced parts, and
system clearance.
100% of reports
submitted to LSU
Biomedical
Engineering and
Respiratory
Therapy within 10
calendar days of
service
completion.
Document
review against
Joint
Commission
standards.
-- 3 of 11 --
Department of Health and Human Services
Indian Health Service
4. CONTRACTOR RESPONSIBILITIES
4.1. Preventative Maintenance and Recertification
The Contractor shall perform one (1) comprehensive annual Preventative Maintenance
inspection and system recertification visit per contract year, scheduled within 30 days
prior to or 30 days after the annual service date.
PM procedures must conform to Medical Graphics Corporation (MGC) guidelines and
include, at a minimum:
Comprehensive physical and mechanical inspection of the body plethysmograph
cabin, door seals, latching mechanisms, and structural components.
Electronic calibration of pressure transducers, pneumotachographs, flow sensors,
gas analyzers, and automated valve circuits.
System leak checks, baseline electronic alignment, and verification of software
diagnostic routines.
Affixing a visible, legible inspection sticker indicating date inspected, next due date,
technician signature, and pass/fail certification.
# Required Service
(Outcome)
Performance
Standard
Acceptable
Quality Level
(AQL)
Method of
Surveillance
3.6
Tech…